Operator‘s Manual for instrument types MP0R 37
Instrument Settings
Data transmission - Menu 4
Set the data to be transmitted during and after the measurement.
Call up instrument
setting menu 4 see description on page 31.
Group separator - Menu 4
The end of a block can be marked with a group separator. The group sepa-
ra
tor can be transferred to the computer together with the readings or
mean values.
Call up instrument setting menu 4 see description on page 31.
Auto switch-off - Menu 4
Use this menu topic to set time period after which the instrument will switch
OF
F automatically if no key is pressed, no measurement is carry out, or no
data are transferred to the PC.
Mode Default setting Further settings
Std
= each single reading
will be transmitted to the PC
= mean values only will
be transmitted to the PC
SSPC Always single readings will be
transmitted
not available - fix instrument
setting, not changeable
90.10 Always mean values will be
transmitted
not available - fix instrument
setting, not changeable
Mode Default setting Further settings
STD,
90.10
= group separator trig-
gering enabled
= group separator trig-
gering disabled
SSPC not available - Triggering of group separator not possible.