Operator‘s Manual for instrument types MP0R 41
Trouble Shooting
10 Trouble Shooting
Message Explanation Solution
Dark
display
Instrument switched OFF. Press the button [OK] to
turn the instrument ON.
Battery contact foil still in
the battery compartment
and the instrument can not
switched ON.
Remove the battery contact
foil before (first) use.
The USB interface is in an
undefined condition.
Remove the batteries from
the instrument and insert
the batteries again. Press
the button [OK] to turn the
instrument ON.
Er1 Internal error Perform a normalization. If
the error occurs repeatedly,
call the Fischer servicing.
www.helmut-fischer.com
Er4 Overflow of the data
memory.
Delete the data memory,
18.
Er5 Cannot determine value for
substrate material.
Button [OK] would be
pressed although no mea-
surement was performed
on the uncoated reference
part (base). This results in
the termination of the nor-
malization or corrective
calibration.
Use parts with applicable
base material, see pages
12 and 14.
Call the desired calibration
menu ( , ) again and
perform 5 to 10 measure-
ments on the uncoated ref-
erence part (base).
O
CAL