Operator‘s Manual for instrument types MP0R 45
Trouble Shooting
10.1 Data transmission via USB - trouble shooting
Problem Explanation Solution
Calibration/
Normaliza-
tion not
possible
The lock function is active
for the calibration or normal-
ization functions (
or
).
Deactivate the lock function
in service menu 1, 33.
measure-
ment
values also
after cali-
bration too
high.
An Offset value is still set. Set the Offset value to zero
22.
Problem Explanation Solution
Erroneous
data trans-
mission via
USB
USB connectors may have
different quality. Especially
the connectors on the PC
front side have problems
with the data transmission.
Use an USB connection on
the rear side of the PC.
Possible that for this (new)
USB connector you have to
install the USB driver again.
23.
Instrument
can not
switched
ON
The USB interface is in an
undefined condition.
Remove the batteries from
the instrument and insert
the batteries again. Press
the button [OK] to switch
the instrument ON.
Impossible
to open the
evaluation,
The data storage is deacti-
vated.
Activate the data storage in
the instrument setting menu
1, 34.