44 Operator‘s Manual for instrument types MP0R
Trouble Shooting
Er28 Internal instrument error:
Not able to compute
measurement.
Different reference parts
with different base materi-
als are used for corrective
calibration.
Use always one reference
part for the corrective cali-
bration.
Base: measurement on
uncoated reference part.
STD: place the calibration
foil on the reference part
used for base measure-
ment. Then perform mea-
surements on the calibra-
tion foil.
Er29 The locked function has
been called (
or ).
Set the function free again
in instrument setting menu
1, 33.
The lock function for the cali-
bration functions (
or
) is active.
Set the function free again
in instrument setting menu
1, 33.
0.00 Instrument setting menu
quitted and no measure-
ment captured yet.
Perform measurements on
the specimen.
- ---- No measurement storage.
Free-Running Display
mode is active.
Value is out of the measu-
ring range.
The USB interface is in an
undefined condition.
End of the list of the single
readings.
Deactivate the mode in the
instrument setting menu 5,
39.
See under Er06, 42.
Remove the batteries from
the instrument and insert
the batteries again. Press
the button [OK] to turn the
instrument ON.
-ECAL No Master Calibration Contact customer service.
Message Explanation Solution
O
CAL
O
CAL