32 Operator‘s Manual for instrument types MP0R
Instrument Settings
9.2 Instrument setting menus
Measurement modes Std, 90.10, SSPC - Menu 1
Additional to the common coating thickness measurement, measurement
mode
s can be select for measurements according to the specifications of
the IMO PSPC (90.10) and of the SSPC-PA2.
Call up instrument setting menu 1 see description on page 31.
Default setting Further settings (use [ ] or [ ])
or
Common coating thickness measurement
Coating thickness measurement according to the "Performance
Standard for Protective Coatings" (PSPC) from the International
Maritime Organization (IMO).
Nominal Dry Film Thickness set at 320 µm ( ).
No block size setting (not available: menu 1, parameter ).
Data storage always enabled (not available: menu 1, parame-
ter )
Only single readings transmission (not available: menu 4, pa-
rameter ).
Coating thickness measuring in compliance with the SSPC-PA2
specification from the Society for Protective Coatings.
Block size 3, default setting (menu 1, parameter )
Data storage always enabled (not available: menu 1, parame-
ter )
Only mean value transmission (not available: menu 4, parame-
ter ).
No group separator transmission (not available: menu 4, pa-
rameter )
TOL