16 Operator‘s Manual for instrument types MP0R
Influencing Variables Taken into Consideration
6. Perform 5 to 10 measurements on the calibration foil. For this purpose,
place the probe inside the circle onto the foil. After each measurement,
the current single reading will appear on the display.
7. Use the arrow key [] or [] to set the nominal value for the calibration
foil, e.g., "71.9
μ
m". The nominal value is printed on the calibration foil.
8. Press button [OK]. The reference point and correction factor received
from the calibration will be stored in the instrument.
The instrument is ready to make measurements.
Please Note!
If you want to delete wrong readings, press button [ESC]. Then, the
complete data memory will be deleted, but the corrective calibration routine
will not be interrupted. After that, perform five more measurements and go
on with the next step.
Please Note!
Only for DUALSCOPE
®
instrument types: If you want to measure on both
substrate materials a Corrective Calibration procedure must be carried out
for each substrate material (Fe and NFe).