Operator‘s Manual for instrument types MP0R 15
Influencing Variables Taken into Consideration
Before you start
Instrument must be turned ON by pressing the key [OK].
For MP0R instrument models only: Remove the orange protective cap
from the probe.
Corrective Calibration procedure
1. Open the calibration menu: Set the display menu with [] or [] to
show
.
2. Press button [OK]. The message "Base" (for “substrate (base) materi-
al”) appears.
3. Perform 5 to 10 measurements on the uncoated reference part (=
Base, specimen without the coating). After each measurement, the
current reading will appear on the displays.
4. Press button [OK]. The value "0.00" and "STD1" (for calibration foil)
appears. The reference point received from the normalization (step #3)
will now be stored in the instrument as zero point.
5. Place the calibration foil on the uncoated specimen.
Please Note!
All readings in the memory will be deleted when performing a corrective cali-
bration.
Please Note!
The menu „Calibration“ can be locked to protect the instrument from unin-
tended interventions. Procedure 31 and 33.
CAL
Calibration foil
The circle on the calibration
foil must be on the marked
measurement area of the
un-coated reference part.
Foil thickness,
nominal value
Guaranteed error
limit
Order no. of the
calibration foil
Uncoated speci-