Operator‘s Manual for instrument types MP0R 29
Data Transmission
3. Set the display menu with [ ] or [ ] to show and press button [OK]
to delete the measurement memory. After this delete command, the in-
strument will sent a group separator to the PC automatically or at next
measurement capture.
Triggering the group separator automatically to the PC
Before you start
USB driver is installed on the computer where the instrument will con-
nected.
The instrument is connected at the computer via the USB cable.
The instrument is turned ON ([OK] key).
For instrument models MP0R only: Remove the orange protective cap
from the probe.
Settings of service menu 1 must be bL = between 2 (mode SSPC: 3)
and 20 (number of readings, which are collected in one block). Proce-
dure 31 and 33.
Settings of service menu 4 (Procedure 31 and 37) must be
SI = 1 or 0 (for mode Std only; 1 = only single readings, 0 = only mean
values are transmitted),
GS = 1 (group separator is active).
The program on the computer to which the data will transmitted is
open. Attend the manual of the program for data import.
Procedure
A group separator is sent automatically at each block creation. The collec-
tion to a block performs automatically the computation of the block statis-
tics and the subsequent deletion of the data of the block in the
measurement memory.
Please Note!
Not available in measuring mode 90.10.
DEL