8 – Specifications
X-series User’s Manual
87
8.5 Performance Characteristics
Power Consumption
FLIR PWR Supply @ 120V
AC
Continuous Normal w/NUC Flag:
Power Consumption
Camera DC Power @ 24V
DC
Continuous Normal w/NUC Flag:
≈7 minutes to reach operating temperature
Sensitivity (w/o optics)
NEΔT
1
InSb: <20mK (18 mK typ.)
SLS: <45mK (40 mK typ.)
Factory Calibration
Temperature Measurement
Ranges
InSb:
-20°C to 350°C (-4°F to 662°F)
45°C to 600°C (113°F to 1,112°F
250°C to 2000°C (482°F to 3,632°F)
500°C to 3,000°C (932°F to 5,432°F)
SLS:
-20°C to 650°C (-4°F to 1,202°F)
250°C to 1,500°C (482°F to 3,632°F)
500°C to 3000°C (932°F to 5,432°F)
Factory Calibration
Temperature Measurement
Accuracy
2
± 1°C or ± 1% of reading
0°C to +3000°C (17mm, 25mm, 50mm, 100mm, 200mm
Lenses)
± 2°C or ± 2% of reading
-20°C to 0°C (17mm, 25mm, 50mm, 100mm, 200mm lenses)
-10°C to +500°C (1X & 4X Closeup lenses)
1) NET is at 50% nominal bucket fill, 298K background, + 5
o
C signal
2) Assuming ideal laboratory conditions (close range, known high emissivity targets, stabilized camera and lens temperatures) and
only if the lens and camera are matched, i.e. not a 3-5 micron lens on a broadband camera.
8.6 Non Uniformity Correction
One Point (offset value with unity gain)
Two Point (offset and gain values, non-volatile)
Two Point w/Bad Pixel Detection/Replacement