Juno System: User Guide
47
Appendix A: Troubleshooting
Powering ON/OFF
Observation Possible Cause Recommended Action
Juno™ system failed to
power on
No AC power Verify that the power cable is properly connected to the
power socket and the instrument.
Verify that the power socket has power.
Verify that the toggle switch on the back of the instrument
is in the on position.
No display after power is
turned on and fan is
running
Juno system in sleep
mode
Touch the screen or push the front power button for
~2 seconds to wake up the instrument. The system shutdown
screen displays, which you can cancel by tapping NO. The
system enters sleep mode when the start screen is displayed
and there is no user input for 1 hour.
No display after power is
turned on and fan is not
running
Juno system power
supply not fully reset
Turn off the Juno system power switch on the back panel.
Wait 10 seconds and reboot by turning on the power switch.
All components power on.
System check error
Unrecoverable error
Error during system
calibration
Reboot the system. If the system fails to recover, then contact
Fluidigm technical support and note any error codes.