Juno System: User Guide
Appendix B: Preventive Maintenance, Decontamination, and Disposal
Decontamination and Disposal
56
Decontamination and Disposal
WARNING BIOHAZARD. If you are putting biohazardous material on the
instrument, use appropriate personal protective equipment and adhere to Biosafety
in Microbiological and Biomedical Laboratories (BMBL), a publication from the
Centers for Disease Control and Prevention, and to your lab's safety protocol to limit
biohazard risks. If biohazardous materials are used, properly label the equipment as
a biohazard. For more information, see the BMBL guidelines online at
cdc.gov/biosafety/publications/index.htm.
Ensure that the Juno system is cleaned and/or decontaminated prior to servicing the
equipment, removing it from use, or transporting it for disposal. Refer to the instructions
contained in this document and use only those materials specified.
Biological Agents
1 Using a soft cloth, apply 70% ethyl alcohol or 70% isopropyl alcohol to all accessible
surfaces.
2 Keep surfaces wet for at least 5 minutes, then wipe dry.
3 Repeat steps 1 and 2 once.
4 Clean all decontaminated surfaces with a wet cloth to remove residual alcohol and wipe
dry.
Hazardous Chemicals
1 Using a soft cloth, apply 70% ethyl alcohol or 70% isopropyl alcohol to all accessible
surfaces.
IMPORTANT Before use, ensure that alcohol is compatible with the chemicals used.
2 Keep surfaces wet for at least 5 minutes, and then wipe dry.
3 Repeat steps 1 and 2 once.
4 Clean all decontaminated surfaces with a wet cloth to remove residual alcohol and wipe
dry.
Radioactive Materials
1 Using a soft cloth, apply an industry standard radioactivity decontaminant to all
accessible surfaces.
2 Wipe the surfaces as directed by the decontaminant manufacturer.
3 Survey the instrument with an appropriate radioactivity measuring device.
4 Ensure that the survey results are at or below background level.