Remote Commands
Commands
6
6-29
OUT
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xx x
(Output command) Sets the output of the Calibrator and establishes a new reference
point for the error mode. If only one amplitude is supplied, the Calibrator sources a
single output. If two amplitudes are supplied, the Calibrator sources two outputs. The
second amplitude will be sourced at the AUX terminals for dual voltage outputs. If the
frequency is not supplied, the Calibrator will use the frequency that is presently in use.
To source or measure a temperature, select the desired sensor and sensor parameters
first. (See the TSENS_TYPE, RTD_*, and TC_* commands.)
To source a signal using the Calibrator scope options, refer to the SCOPE command in
Chapter 8.
If you change the frequency of an ac function and the harmonic output is not explicitly
set at the same time with the HARMONIC command, the harmonic will be set to 1.
Use multipliers e.g., k, M, µ with the OUT command, as desired.
Parameters: <value> V Volts dc or update volts ac
<value> DBM Volts ac dBm update
<value> V, <value> Hz Volts ac or volts dc with 0 Hz
<value> DBM, <value> Hz Volts ac in dBm
<value> A Current dc or update current ac
<value> A, <value> Hz Current ac
<value> OHM Resistance
<value> F Capacitance
<value> CEL Temperature (Celsius)
<value> FAR Temperature (Fahrenheit)
<value> HZ Update frequency
<value> V, <value> A Power dc or update power ac
<value> V, <value> A, <value> HZ Power ac
<value> V, <value> V Dual volts dc or update dual ac
<value> V, <value> V, <value> HZ Dual volts ac in volts
<value> For single output, changes
amplitude keeping unit and
frequency the same.