5520A
Operators Manual
6-30
Examples: OUT 15.2 V (volts; 15.2 V @ same frequency)
OUT 20 DBM (volts; 20 dBm @ same frequency)
OUT 10 V, 60 Hz (volts ac; 10 V @ 60 Hz)
OUT 10 DBM, 50 HZ (volts ac; 10 dBm @ 50 Hz)
OUT 1.2 MA (current; 1.2 mA @ same frequency)
OUT 1 A, 400 HZ (current ac; 1 A @ 400 Hz)
OUT 1 KOHM (ohms; 1 kΩ)
OUT 1 UF (capacitance; 1µF)
OUT 100 CEL (temperature; 100 °C)
OUT −32 FAR (temperature; −32°F)
OUT 60 HZ (frequency update; 60 Hz)
OUT 10 V, 1 A (power; 10 watts @ same frequency)
OUT 1 V, 1 A, 60 HZ (power ac; 1 watts @ 60 Hz)
OUT 1 V, 2 V (dual volts; 1 V, 2 V @ same freq.)
OUT 10 MV, 20 MV, 60 HZ (dual volts; .01 V, .02 V @ 60 Hz)
Each example shows a value and unit, e.g., −15.2 V. If a value is entered without a unit,
the value of the existing output is changed, when logically allowed.
OUT?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Output query) Returns the output amplitudes and frequency of the Calibrator.
Multipliers (e.g., K or M) are not used in the response.
Parameters: V (optional for ac voltage and TC modes)
DBM (optional for ac voltage modes)
CEL (optional for RTD and TC modes, Celsius)
FAR (optional for RTD and TC modes, Fahrenheit)
OHM (optional for RTD modes, ohms)
Response: <primary amplitude value>,<primary units>,
<secondary amplitude value>,<secondary units>,
<fundamental frequency value>
Examples: OUT? returns −1.520000E+01,V,0E+00,0,0.00E+00
OUT? returns 1.88300E−01,A,0E+00,0,4.420E+02
OUT? returns 1.23000E+00,V,2.34000E+00,V,6.000E+01
OUT? returns 1.92400E+06,OHM,0E+00,0,0.00E+00
OUT? returns 1.52000E+01,V,1.88300E−01,A,4.420E+02
OUT? DBM returns 2.586E+01,DBM,0E+00,A,4.420E+02
OUT? returns 1.0430E+02,CEL,0E+00,0,0.00E+00
OUT? FAR returns 2.19740000E+02,FAR,0E+00,0,0.00E+00
OUT? V returns 4.2740E−03,V,0E+00,0,0.00E+00
OUT? OHM returns 1.40135E+02,OHM,0E+00,0,0.00E+00