The respective values for the above examples are:
15.2 V
188.3 mA, 442 Hz
1.23 V, 2.34 V, 60 Hz
1.924 M
15.2 V, 188.3 mA, 442 Hz
25.86 dBm, 442 Hz (25.86 dBm = 15.2 V at 600 )
104.3 °C
219.74 °F (same value as 104.3 °C, in Fahrenheit)
4.274 mV (same value as 104.3 °C for a K-type thermocouple, in volts)
140.135 (same value as 104.3 °C for a pt385 RTD, in ohms)
The primary and secondary units are: V, DBM, A, OHM, F, CEL, FAR. The units for the
<frequency value> is always assumed to be Hz.
OUT_ERR?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Output Error query) Returns the UUT error and units computed by the Calibrator after
shifting the output with the INCR command. The return units are PPM (parts per
million), PCT (percent), DB (decibels) or 0 if there is no error. The UUT error is not
computed when editing frequency.
Response: <value of error>,<units>
Example: OUT_ERR? returns 1.00000E+01,PCT
Returns 10% when the UUT is reading low by 10%.
PHASE
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xxx
(Phase Difference command) Sets a phase difference between the Calibrator front panel
NORMAL and AUX or 20A terminals for ac power and ac dual voltage outputs. The
NORMAL terminal output is the phase reference. The set range is 0.00 to ±180.00
degrees, with + for a leading phase difference and for a lagging phase difference.
Parameter: <phase value> DEG (DEG, for degree, is optional)
Example: PHASE 60 DEG
Set the phase difference so the frequency output at the AUX terminals lags the frequency
output at the NORMAL terminals by 60 degrees.
PHASE?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Phase Difference query) Returns the phase difference between the Calibrator front
panel NORMAL and AUX terminals for ac power and ac dual voltage outputs.
Response: <phase value>