6-32
Example: PHASE? returns 6.000E+01
Returns 60 when the frequency output at the AUX terminals is lagging the frequency
output at the NORMAL terminals by 60 degrees.
POWER?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Calculate Power Output query) Returns the equivalent real power for ac and dc power
outputs, based on the voltage and current settings, and power factor (ac only). If the
output is not ac or dc power, the return is 0E+00 (zero) watts.
Response: <value> (in watts)
Example: POWER? returns 1.00000E+01
Returns 10 when the output voltage is 10 V dc and output current 1 A dc, for 10 watts
real power.
Example: POWER? returns 1.00000E+01
Returns 10 when the output voltage is 10 V ac and output current 2 A ac and power
factor is .5, for 10 watts real power.
PR_PRT X IEEE X RS=232
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
Description: Prints a self calibration report out the selected serial port.
Parameters: 1. Type of report to print: STORED, ACTIVE, or CONSTS
2. Format of report: PRINT (designed to be read), SPREAD
(designed to be loaded into a
spreadsheet )
3. Calibration interval to be used for instrument specifications in the
report: I90D (90 day spec), I1Y (I year spec)
4. Serial port through which to print: HOST, UUT
Example: PR-PRT STORED, PRINT, I90D, HOST
PRES?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Pressure Module query) Queries the attached pressure module for its model and serial
number.
Responses: (Indefinite ASCII) A message containing four fields separated by commas
as follows:
1. Manufacturer
2. Model number
3. Serial number
4. Firmware revision (0)
Example: FLUKE,700P05,9467502,0