Maintenance
Performance Tests
7
7-17
Table 7-8. Verification Tests for AC Current (cont)
Range Output Frequency Lower Limit Upper Limit
2.99999 A 2.99000 A 10 kHz 2.92520 A 3.05480 A
20.5000 A 3.3000 A 500 Hz 3.2954 A 3.3046 A
20.5000 A 3.3000 A 1 kHz 3.2954 A 3.3046 A
20.5000 A 3.3000 A 5 kHz 3.2155 A 3.3845 A
20.5000 A 10.9000 A 45 Hz 10.8926 A 10.9075 A
20.5000 A 10.9000 A 65 Hz 10.8926 A 10.9075 A
20.5000 A 10.9000 A 500 Hz 10.8893 A 10.9107 A
20.5000 A 10.9000 A 1 kHz 10.8893 A 10.9107 A
20.5000 A 10.9000 A 5 kHz 10.6255 A 11.1745 A
20.5000 A 20.0000 A 45 Hz 19.9750 A 20.0250 A
20.5000 A 20.0000 A 65 Hz 19.9750 A 20.0250 A
20.5000 A 20.0000 A 500 Hz 19.9690 A 20.0310 A
20.5000 A 20.0000 A 1 kHz 19.9690 A 20.0310 A
20.5000 A 20.0000 A 5 kHz 19.4950 A 20.5050 A