5520A
Operators Manual
7-18
Table 7-9. Verification Tests for Capacitance
Range Output
Test
Frequency
or Current Lower Limit Upper Limit
0.3999 nF 0.1900 nF 5 kHz 0.1793 nF 0.2007 nF
0.3999 nF 0.3500 nF 1 kHz 0.3387 nF 0.3613 nF
1.0999 nF 0.4800 nF 1 kHz 0.4682 nF 0.4918 nF
1.0999 nF 0.6000 nF 1 kHz 0.5877 nF 0.6123 nF
1.0999 nF 1.0000 nF 1 kHz 0.9862 nF 1.0138 nF
3.2999 nF 2.0000 nF 1 kHz 1.9824 nF 2.0176 nF
10.9999 nF 7.0000 nF 1 kHz 6.9767 nF 7.0233 nF
10.9999 nF 10.9000 nF 1 kHz 10.8693 nF 10.9307 nF
32.9999 nF 20.0000 nF 1 kHz 19.8620 nF 20.1380 nF
109.999 nF 70.000 nF 1 kHz 69.767 nF 70.233 nF
109.999 nF 109.000 nF 1 kHz 108.693 nF 109.307 nF
329.999 nF 200.000 nF 1 kHz 199.320 nF 200.680 nF
329.999 nF 300.000 nF 1 kHz 299.130 nF 300.870 nF
1.09999 µF 0.70000 µF 100 Hz 0.69767 µF 0.70233 µF
1.09999 µF 1.09000 µF 100 Hz 1.08693 µF 1.09307 µF
3.29999 µF 2.00000 µF 100 Hz 1.99320 µF 2.00680 µF
3.29999 µF 3.00000 µF 100 Hz 2.99130 µF 3.00870 µF
10.9999 µF 7.0000 µF 100 Hz 6.9767 µF 7.0233 µF
10.9999 µF 10.9000 µF 100 Hz 10.8693 µF 10.9307 µF
32.9999 µF 20.0000 µF 100 Hz 19.9100 µF 20.0900 µF
32.9999 µF 30.0000 µF 100 Hz 29.8800 µF 30.1200 µF
109.999 µF 70.000 µF 50 Hz 69.662 µF 70.338 µF
109.999 µF 109.000 µF 50 Hz 108.529 µF 109.471 µF
329.999 µF 200.000 µF 54 µA dc 199.020 µF 200.980 µF
329.999 µF 300.000 µF 80 µA dc 298.680 µF 301.320 µF
1.09999 mF 0.33000 mF 90 µA dc 0.32788 mF 0.33212 mF
1.09999 mF 0.70000 mF 180 µA dc 0.69662 mF 0.70338 mF
1.09999 mF 1.09000 mF 270 µA dc 1.08529 mF 1.09471 mF
3.2999 mF 1.1000 mF 270 µA dc 1.0933 mF 1.1067 mF
3.2999 mF 2.0000 mF 540 µA dc 1.9902 mF 2.0098 mF
3.2999 mF 3.0000 mF 800 µA dc 2.9868 mF 3.0132 mF