Maintenance
Performance Tests
7
7-19
Table 7-9. Verification Tests for Capacitance (cont)
Range Output
Test
Frequency
or Current Lower Limit Upper Limit
10.9999 mF 3.3000 mF 900 µA dc 3.2788 mF 3.3212 mF
10.9999 mF 10.9000 mF 2.7 mA dc 10.8529 mF 10.9471 mF
32.9999 mF 20.0000 mF 5.4 mA dc 19.8300 mF 20.1700 mF
32.9999 mF 30.0000 mF 8.0 mA dc 29.7600 mF 30.2400 mF
110.000 mF 33.000 mF 9.0 mA dc 32.570 mF 33.430 mF
110.000 mF 110.000 mF 27.0 mA dc 108.800 mF 111.200 mF
Table 7-10. Verification Tests for Thermocouple Simulation
TC Type Output, °C Lower Limit, mV Upper Limit, mV
10 µV/°C 0.00 °C (0.0000 mV) -0.0030 0.0030
100.00 °C (1.0000 mV) 0.99696 1.00304
-100.00 °C (-1.0000 mV) -1.00304 -.99696
1000.00 °C (10.0000 mV) 9.99660 10.00340
-1000.00 °C (-10.0000 mV) -10.0034 -9.9966
10000.00 °C (100.0000 mV) 99.9930 100.0070
-10000.00 °C (-100.0000 mV) -100.0070 -99.9930
Table 7-11. Verification Tests for Thermocouple Measurement
TC Type Input, mV Lower Limit, °C Upper Limit, °C
10 µV/°C 0.00 °C (0.0000 mV) -0.30 0.30
10000.00 °C (100.0000 mV) 9999.30 10000.70
-10000.00 °C (-100.0000 mV) -10000.70 -9999.30
30000.00 °C (300.0000 mV) 29998.50 30001.50
-30000.00 °C (-300.0000 mV) -30001.50 -29998.50

Table of Contents