Maintenance
Performance Tests
7
7-21
Table 7-13. Verification Tests for Phase Accuracy, V and I
Range,
Normal
Output
Output,
Normal
Freq-
uency
Range, AUX
Output Output, AUX
Phase
°
Lower
Limit
°
Upper
Limit
°
329.999 mV 30.000 mV 65 Hz 329.99 mA 300.00 mA 0 -0.10 0.10
329.999 mV 30.000 mV 1 kHz 329.99 mA 300.00 mA 0 -0.50 0.50
329.999 mV 30.000 mV 30 kHz 329.99 mA 300.00 mA 0 -10.00 10.00
329.999 mV 200.000 mV 65 Hz 2.99999 A 2.00000 A 0 -0.10 0.10
329.999 mV 50.000 mV 65 Hz 20.5000 A 5.0000 A 0 -0.10 0.10
329.999 mV 50.000 mV 400 Hz 20.5000 A 5.0000 A 0 -0.25 0.25
329.999 mV 30.000 mV 65 Hz 329.99 mA 300.00 mA 60 59.90 60.10
329.999 mV 200.000 mV 65 Hz 2.99999 A 2.00000 A 60 59.90 60.10
329.999 mV 200.000 mV 65 Hz 20.5000 A 20.0000 A 60 59.90 60.10
329.999 mV 200.000 mV 400 Hz 20.5000 A 20.0000 A 60 59.75 60.25
32.9999 V 3.3000 V 65 Hz 329.99 mA 300.00 mA 0 -0.10 0.10
32.9999 V 3.3000 V 65 Hz 2.99999 A 2.00000 A 0 -0.10 0.10
32.9999 V 3.3000 V 65 Hz 20.5000 A 5.0000 A 0 -0.10 0.10
32.9999 V 3.3000 V 400 Hz 20.5000 A 5.0000 A 0 -0.25 0.25
32.9999 V 3.3000 V 65 Hz 329.99 mA 300.00 mA 90 89.90 90.10
32.9999 V 3.3000 V 65 Hz 2.99999 A 2.00000 A 90 89.90 90.10
32.9999 V 3.3000 V 65 Hz 20.5000 A 20.0000 A 90 89.90 90.10
32.9999 V 3.3000 V 400 Hz 20.5000 A 20.0000 A 90 89.75 90.25
329.999 V 33.000 V 65 Hz 329.99 mA 300.00 mA 0 -0.10 0.10
329.999 V 33.000 V 65 Hz 2.99999 A 2.00000 A 0 -0.10 0.100
329.999 V 33.000 V 65 Hz 20.5000 A 5.0000 A 0 -0.10 0.10
329.999 V 33.000 V 400 Hz 20.5000 A 5.0000 A 0 -0.25 0.25
329.999 V 33.000 V 65 Hz 329.99 mA 300.00 mA 90 89.90 90.10
329.999 V 33.000 V 65 Hz 2.99999 A 2.00000 A 90 89.90 90.10
329.999 V 33.000 V 65 Hz 20.5000 A 20.0000 A 90 89.90 90.10
329.999 V 33.000 V 400 Hz 20.5000 A 20.0000 A 90 89.75 90.25