5520A
Operators Manual
7-22
Table 7-14. Verification Tests for Frequency
Range, Normal
Output, V
Output,
Normal, V Frequency
Lower Limit
(Note)
Upper Limit
(Note)
3.29999 3.00000 119.00 Hz 118.99970 Hz 119.00030Hz
120.0 Hz 119.99970 Hz 120.00031 Hz
1000.0 Hz 999.9975 Hz 1000.0025 Hz
100.00 kHz 99,999.75 Hz 100,000.25 Hz
Note: Frequency accuracy is specified for 1 year