EasyManua.ls Logo

Fluke 5520A - Page 296

Fluke 5520A
420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5520A
Operators Manual
8-12
8-14. Oscilloscope Input Resistance Measurement Specifications
Table 8-12. Oscilloscope Input Resistance Measurement Specifications
Scope input selected 50 1 M
Measurement Range 40 to 60 500 k to 1.5 M
Uncertainty 0.1 % 0.1 %
8-15. Oscilloscope Input Capacitance Measurement Specifications
Table 8-13. Oscilloscope Input Capacitance Measurement Specifications
Scope input selected 1 M
Measurement Range 5 pF to 50 pF
Uncertainty ± (5% of input + 0.5 pF) [1]
[1] Measurement made within 30 minutes of capacitance zero reference. Scope option must be selected
for at least five minutes prior to any capacitance measurement, including the zero process.
8-16. Overload Measurement Specifications
Table 8-14. Overload Measurement Specifications
Source
Voltage
Typical ‘On’ current
indication
Typical ‘Off’ current
indication
Maximum Time Limit DC
or AC (1 kHz)
5 V to 9 V 100 mA to 180 mA 10 mA setable 1s to 60s

Table of Contents

Other manuals for Fluke 5520A

Related product manuals