724/725/726
Calibration Manual
12
General Specifications
Operating temperature
724 and 725
726
-10 °C to 55 °C
-10 °C to 50 °C
Storage temperature - 20 °C to 70 °C
Stability ± 0.005 % of range/°C outside of 23 ± 5 °C
Operating altitude 3000 meters above mean sea level
Relative Humidity (% RH operating without
condensation)
90 % (10 to 30 °C)
75 % (30 to 40 °C)
45 % (40 to 50 °C)
35 % (50 to 55 °C)
uncontrolled < 10 °C
Vibration Random, 2 g, 5 Hz to 500 Hz
Safety
724 and 725
726
EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA
C22.2 No 1010.1:1992
EN50082-1:1992 and EN55022: 1994 Class B Criteria
A or B CSA C22.2 No 1010.1:1992
Power requirements 4 AA alkaline batteries
Protection Class Pollution Degree II
Size 96 x 200 x 47 mm (3.75 x 7.9 x 1.86 in)
Weight 650 gm (1 lb, 7 oz)
Cleaning the Calibrator
XWWarning
To avoid electric shock or damage to the Calibrator, never allow water
inside the Calibrator’s case.
If the Calibrator requires cleaning, wipe it down with a cloth that is lightly dampened
with water or a mild detergent.
Do not use aromatic hydrocarbons, chlorinated solvents, or methanol-based fluids when
cleaning the Calibrator. To avoid damaging the case, never apply solvents to the
Calibrator.
Replacing the Batteries
XWWarning
To avoid electric shock, remove the test leads from the Calibrator
before opening the battery door. Close and latch the battery door
before using the Calibrator.
To avoid false readings, which can lead to possible electric shock or
personal injury, replace the batteries as soon as the low battery
indicator (
B) appears.
When replacing the batteries, always use four new AA batteries. Never
mix new and used batteries in the Calibrator.