EasyManua.ls Logo

Fluke 724 - Fluke 724 Calibrator Maintenance; Cleaning the Fluke 724 Calibrator; Service Center Calibration and Repair

Fluke 724
76 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
724/725/726
Calibration Manual
38
Lower Display Ohms Source Tests
1. Press F ( R on the 726) on the UUT until appears on the lower display. If
necessary, press M until SOURCE appears on the lower display.
2. Set the Fluke 8508A to measure 4-wire resistance.
3. Set the Fluke 8508A to 2 k range for 72x resistance < 400 , to 20 k range for
400 or more.
4. Make the connections shown in Figure 12.
5. Use the arrow keys on the UUT to set the UUT output to the resistance values in
Table 18 and verify that the Fluke 8508A readings are within the limits shown.
aal13f.eps
Figure 12. Lower Display Ohms Source Test Connections
Table 18. Lower Display Ohms Source Readings
UUT Settings
724/725
Fluke 8508A Readings
726
Fluke 8508A Readings
Range
15.0 14.9 to 15.1 14.94775 to 15.05225 2 k
360.0 359.9 to 360.1 359.896 to 360.104 2 k
500 499.5 to 500.5 499.375 to 500.625 20 k
1500 1499.5 to 1500.5 1499.275 to 1500.725 20 k
3200 3199.0 to 3201.0 - 20 k
3800 - 3798.93 to 3807.01 20 k
The performance tests for the 724 are now complete. Disconnect and secure all test
equipment.

Table of Contents

Other manuals for Fluke 724

Related product manuals