EasyManuals Logo

Fluke 724 Calibration Manual

Fluke 724
76 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #32 background imageLoading...
Page #32 background image
724/725/726
Calibration Manual
24
Performance Tests
XXWarning
Some of the performance tests involve the use of high voltages and
should be performed by qualified personnel only.
To avoid electric shock, always set the calibration source (5520A) to
Standby (STBY) mode between tests and before handling the test
connections or test cables.
Performance tests confirm the complete operability of the Calibrator and check the
accuracy of each function against the Calibrator’s specifications.
Note
If the Calibrator fails any performance test, it needs calibration
adjustments. If the Calibrator continues to fail after calibration
adjustments, send it to an authorized Fluke Service Center for repair.
The Calibrator’s performance and accuracy are specified for one year after calibration at
operating temperatures of +18 °C to +28 °C (64 °F to 82 °F), in relative humidity to
90 %. The specifications assume the Calibrator has been warmed up for five minutes
before use.
It is not necessary to open the case for the performance tests; no mechanical adjustments
are necessary. Merely make the required connections, source the designated values, and
determine if the reading on the Calibrator or the multimeter falls within the acceptable
range indicated.
Preparing for the Performance Tests
Notes
Performance tests for the Calibrator can be performed manually, or they
can be computer-automated (using Fluke Metrology Software). The
Metrology Software automates all of the performance verification tasks,
except for connection of the standards to the Calibrator. This document
provides the procedures necessary for manual performance tests. Contact
Fluke for information on Metrology Software.
These performance tests assume that the person performing them knows
how to use the Calibrator and the required equipment.
Do not attempt to perform these tests unless you are qualified to do so.
Throughout the performance tests, “UUT” (unit under test) refers to the
Calibrator; the word “multimeter” is reserved for the digital multimeter
identified in the required equipment listed in Table 6.
Unless otherwise indicated, all connection diagrams for the verification
tests in this manual showing a calibrator or digital multimeter use a Fluke
5520A Calibrator or Fluke 8508A DMM. If you are using a different DMM,
make the connections appropriate for your instrument.
To prepare the UUT for the performance tests, do the following:
1. Make sure that you have the required equipment available, see Table 6.
2. Make sure the UUT has fresh batteries. See “Replacing the Batteries” earlier in this
manual.
3. Warm up the multimeter as required by its specifications.
4. Remove all test leads from the front of the UUT.
5. Make sure that the UUT is in a stable ambient temperature between 18 °C and 28 °C
(64.4 °F and 82.4 °F) and that it has been warmed up for five minutes.

Table of Contents

Other manuals for Fluke 724

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 724 and is the answer not in the manual?

Fluke 724 Specifications

General IconGeneral
Temperature Range-10°C to 55°C
Thermocouple TypesJ, K, T, E, R, S, B, L, U, N
Current Measurement0 to 24 mA
Operating Temperature-10°C to 55°C
FunctionProcess calibrator
RTD TypesPt100, Pt500, Pt1000
Voltage Measurement0 to 10 V
Loop Power24 V DC
DisplayDual Display
Power Supply4 AA batteries
Resolution0.1 °C
Storage Temperature-20 °C to 60 °C
Battery Life25 hours
Relative Humidity0 % to 90 %
Safety ComplianceCSA C22.2 No. 1010.1:1992

Related product manuals