Cal ISO Volts
Connect the UUT as shown in Figure 22.
aal03f.eps
Figure 22. ISO Volts Calibration Connections
From the Calibration Menu, enter the cal step for Cal ISO Volts. The PC displays:
Enter 0 Volts - press space bar to continue
Set the Fluke 5520A to output 0.0000 V, then press the space bar on the PC. After a short
while the PC displays the following a calibration constant and new prompt:
Offset = -324
Enter 30.00 Volts - press space bar to continue
Set the Fluke 5520A to output 30.0000 V, then press the space bar on the PC. After a
short while, the PC displays the following calibration constants and prompt:
diff = (Counts - Offset)
30700 = 30376 - -324
Volts per count = 0.000977
- press space bar to continue
Press the space bar to return to the Calibration Menu.

Table of Contents

Other manuals for Fluke 724

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 724 and is the answer not in the manual?

Fluke 724 Specifications

General IconGeneral
Temperature Range-10°C to 55°C
Thermocouple TypesJ, K, T, E, R, S, B, L, U, N
Current Measurement0 to 24 mA
Operating Temperature-10°C to 55°C
FunctionProcess calibrator
RTD TypesPt100, Pt500, Pt1000
Voltage Measurement0 to 10 V
Loop Power24 V DC
DisplayDual Display
Power Supply4 AA batteries
Resolution0.1 °C
Storage Temperature-20 °C to 60 °C
Battery Life25 hours
Relative Humidity0 % to 90 %
Safety ComplianceCSA C22.2 No. 1010.1:1992