724/725/726
Calibration Manual
58
Calibrating mA Output
Connect the UUT as shown in Figure 27.
aal12f.eps
Figure 27. mA Output Calibration Connections
After making the proper connections, type “4” then press the Enter key on the PC
keyboard. Input directions are displayed at the bottom of the calibration menu:
Enter Selection:
First Calibration Point. Enter mA displayed:
Second Calibration Point. Enter mA displayed:
Use the 8508A to measure the two calibration points. Enter each of the two calibration
points as prompted.