135
Prism
Particulars Type: 90° roof prism
Accuracy (avg): > 99%
Environmental protection: IP 67 (submersible, dustproof)
Storage temperature: -20°C to 80°C [-4°F to 176°F]
Operating temperature: -20°C to 60°C [-4°F to 140°F]
Dimensions: approx. 100 x 41 x 35 mm [4” x 1 5/8” x 1 3/8”]
Weight: approx. 65 g [2 1/2 oz.]
Wireless module
Particulars Class 1 connectivity, transmitting power 100 mW
Transmission distance: 10 m [33 ft.]
Complies with FCC rules part 15.247
LED indicators: 1 LED for Bluetooth communication,
3 green LEDs for battery status
Power supply: Batteries 2 x 1.5 V IEC LR6 (“AA“)
Operating time: 14 hours typical use (based upon an operating cycle
of 50% measurement, 50% standby)
Storage temperature: -20°C to 60°C [-4°F to 140°F]
Operating temperature: -10°C to 50°C [14°F to 122°F]
Environmental protection:IP 65 (dustproof and water spray resistant),
shockproof
Dimensions: Approx. 81 x 41 x 34 mm [3 1/8” x 1 11/16” x 1 5/16”]
Weight: Approx. 133 g [4.7 oz.] including batteries and cable
Safety
Electrical IEC 61010-1
Battery IEC 62133
Laser IEC 60825-1, 21 CFR 1040.10, .11 with Laser Notice 50
Electromagnetic Environment
Particulars IEC 61326-1: Basic
Radio Frequency Emissions
Particulars IEC CISPR 11: Group 1, Class
A.
Group 1 have intentionally generated and/or use conductively coupled
radio-frequency energy which is necessary for the internal functioning
of the equipment itself.
Class A equipment is suitable for use in non-domestic locations and/or
directly connected to a low-voltage power supply network.
Carrying case
Particulars Case dimensions: approx. 560 x 330 x 130 mm [22 3/64” x 13” x 7 1/8”]
Weight, including all standard parts: approx. 4.8 kg [10.6 lb]
Appendix