80
MANTIS User's Guide | Chapter 9: Dispensing With the MANTIS
Dispense Settings
See the table below for detailed explanations of each option found in the Dispense area.
Dispense Summary Settings
See the table below for detailed explanations of each option found in the Dispense
Summary area.
Hand Settings
See the table below for detailed explanations of each option found in the Hand area.
Option Default Setting Description
Disable Pump for
Long Delay
True
Enables MANTIS to automatically turn off the
pump for long delay. Set this option to False to
disable this feature.
First Dispense
Diaphragm
Small
Determines which diaphragm is to be dispensed
first.
Update Sequence
Editor During
Dispense
False
Set this option to True to enable updates to the
Sequence Editor during dispense.
Use Dispense
Visualization
True
Set this option to False to deactivate the
dispense indicator.
Waste Station
Location
Default
Default: The waste station is in the default
location. See Hardware Overview on page 10.
Adapter: The waste station is set at the back of
MANTIS (for LC3 only). See Setting Up the LC3
for a Dispense on page 91.
Option Description
Theta Motor 1
Offset
Allows you to realign the theta motor 1 position in case its position
goes wrong when homing. To adjust the value, press Alt+F2 on
the keyboard, then set it manually via the Theta Motor Control
Pad. The up arrow increases the value, and the down arrow
decreases it. The speed slider determines how fast the changes
will occur.
Theta Motor 2
Offset
Allows you to realign the theta motor 2 position in case its position
goes wrong when homing. To adjust the value, press Alt+F2 on
the keyboard, then set it manually via the Theta Motor Control
Pad. The up arrow increases the value, and the down arrow
decreases it. The speed slider determines how fast the changes
will occur.
Option Default Setting Description
Enable Dispense
Summary
False
Set this option to True to enable the MANTIS
software to create a dispense summary log file.
This file will be created once per day as long as
this option is enabled.

Table of Contents