Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
Page 6-21
9. Select OK button to close the screen.
10. Repeat steps one through nine for each Storage and Commitment server on the network. You can edit
or delete server by clicking the relevant button. When you have finished configuring all servers, select
the Exit button to close the Storage and Commitment Screen.
Setup Query & Retrieve
Query & Retrieve Server, are for retrieving the DICOM image information from remote server.
To configure the Query & Retrieve server:
1. Select Query & Retrieve button from the setup screen.
2. Setup Local Setting.
Local Listening Port
Time Out: The time in seconds that the Workstation waits for data sent back by DICOM Retrieve Server.
Days to Cache Retrieved Images
3. Click Add button to display the Query & Retrieve Config screen. Obtain the following information about
the Query & Retrieve server from your network administrator and enter it on this screen:
Alias: Unique name, internal to WorkStation, used to identify the server in the list of available destination
devices.
AE Title: DICOM Application Entity Title.