Preparing for Setup
Invenia ABUS 2.0 – System Setup and Basic Service Manual 3-41
4700-0043-00 Rev. 4
Setup Warnings and Cautions (continued)
The following table describes guidelines for reaching operational
temperatures from storage or transport temperatures.
Electrical specifications
Verification of the system’s voltage setting
Verify that the mains voltage specified for the Invenia ABUS 2.0
is available on-site.
The voltage setting for the unit is found on a label near the
Mains Power Switch on the rear of the system. Refer to the
Invenia ABUS 2.0 Basic User Manual, part number
4700-0041-00, for more information.
Do not operate this unit unless all board covers and frame
panels are securely in place. System performance and cooling
require this.
If the Ultrasound system is very cold or hot, do not turn on its
power until it has had a chance to acclimate to its operating
environment.
Operator Manual(s)
The User Manual(s) should be fully read and understood
before operating the Invenia ABUS 2.0 and kept near the
Ultrasound system for quick reference.
Connecting an Invenia ABUS 2.0 to the wrong voltage level will
damage it.