Replacing the Power Cord
Invenia ABUS 2.0 – System Setup and Basic Service Manual 8-13
4700-0043-00 Rev. 4
Replacing the Power Cord
To remove the power cord,
1. Locate the power cord receptacle on the back of the system.
Figure 8-1. Power Cord in Receptacle
2. Pull the steel retaining clip upwards.
3. Disconnect the power cord from its receptacle by pulling the
power cord straight out.
To replace the power cord
1. Locate the power cord receptacle on the back of the system.
2. Align the power cord prongs to match the receptacle, then
securely push the cord straight back into the receptacle.
3. Pull the steel retaining clip down to retain the plug in the
system.
CAUTION
Removing the power cord without lifting the steel retaining clip
may damage the system.