System Setup
3-2 Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4
Setup Time and Warnings
Average setup time
Unpacking the Invenia ABUS 2.0: 1 hour
DICOM Network Configuration: 3 hours or more, depending
on the configuration
Setup warnings
DANGER
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE
OF OPENING THE AC GROUND LINE (I.G., METER’S
GROUND SWITCH IS OPEN), DON’T TOUCH THE
ULTRASOUND SYSTEM!
CAUTION
To prevent electrical shock, connect the unit to a properly
grounded power outlet. Do not use a three to two prong
adapter. This defeats safety grounding.
CAUTION
Do not wear the ESD wrist strap when you work on live circuits
and more than 30 V peak is present.
CAUTION
Do not operate this unit unless all board covers and frame
panels are securely in place. System performance and cooling
require this.
CAUTION
The system must be supplied by an adequately rated electrical
circuit. The capacity of the supply circuit must be as specified.