Setting up the Invenia ABUS 2.0
Invenia ABUS 2.0 – System Setup and Basic Service Manual 3-47
4700-0043-00 Rev. 4
Calibrating the Touch Screen Display
NOTE: The Touch Screen Display is calibrated from the factory.
Use this procedure only if calibration is noticeably offset.
1. Tap the Configuration icon to display the General
configuration screen and tap the System tab.
2. Tap the Calibrate Screen button to open the calibration
function.
Figure 3-7. Calibrate Screen in System Configuration
3. Follow the onscreen instructions or tap the X in the upper
right corner to close the configuration screen.