Page #242 background imageLoading...
Page #242 background image
Diagnostics/Troubleshooting
7-48 Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4
Error Code List
Error
Code
Category Description
11001 Acquisition
Failed to initialize acquisition (missing frontend JSON file,
possible transducer disconnection)
12001 Acq-scanhead Scanhead id does not match.
13001 Renderer Image renderer initialization failed.
14001 System status GPU Max temperature exceeded
14002 System status Power Max exceeded
21001 ScannerApp Init Initialization success
21002 ScannerApp Init Weight motor init home failed
21003 ScannerApp Init Scan motor init home failed
21004 ScannerApp Init Ball motor init home failed
21005 ScannerApp Init Scanhead controller USB COM port not detected
21006 ScannerApp Init Weight controller USB COM port not detected
21007 ScannerApp Init Scanhead and Weight controller USB COM port not detected
22001 ScannerApp USB Scanhead controller USB COM port not detected
22002 ScannerApp USB Weight controller USB COM port not detected
31000 GEUS Unknown Unknown GEUS Error Code received
31001 GEUS Failure GEUS fatal error requiring system shutdown
31002 GEUS Failure GEUS fatal error requiring system restart
31001 GEUS Failure GEUS fatal error DSP not detected requiring system restart
32001 GEUS Info GEUS information error
32002 GEUS Info GEUS information error requiring system restart
33001 GEUS Service GEUS service type error information category

Table of Contents