Invenia ABUS 2.0 Error Log
Invenia ABUS 2.0 – System Setup and Basic Service Manual 7-47
4700-0043-00 Rev. 4
Ultrasound Box (GEUS) error code examples: (continued)
Example 2
Error code 35007 is related to a problem with the Ultrasound
Box fans (fan tray assembly). This problem could be caused by
a a defective fan in the fan tray assembly, or an issue with 12V
fan voltage supplied by MFESP board.
The service team can easily search the Service Manual by the
error code and follow the troubleshooting steps: check the
MFESP fan monitoring readings and run the system diagnostics
available in the Service Desktop to determine the root cause of
the problem.
The Invenia 2.0 system displays an error/information message
along with an error code, and codes are linked to
troubleshooting action items in the Service Manual. However,
some error codes/problems may still require the service team to
collect and to analyze the system error logs.