Functional Checks
4-14 Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4
Mechanical Functions Checks
Scan Head Assembly Functional Checks
Ensure that the Scan Head Assembly:
Adjusts the Image Orientation to the Scan View being
performed.
Compression buttons work as specified.
Start button initiates the scan.
Stop button stops the scan.
Rotates 360 degrees.
Arm locks into position after compression has been applied.
Articulating Arm Functional Checks
Ensure that the Articulating Arm:
Can be repositioned easily.
Can be moved up/down with ease.
Can be locked into position for transport.
Can support the Scan Head Assembly.
Touch Screen Monitor Functional Checks
Ensure that the Touch Screen Monitor:
Can be used to adjust the Scan View workflow.
Can be used to specify Scan Depth.
Can be used to type in Patient information.
Doesn’t drop any pixels.
Casters (Wheels) and Brakes Functional Checks
Examine the wheels frequently for defects to avoid breaking or
jamming and the brakes lock.