Invenia ABUS 2.0 Error Log
Invenia ABUS 2.0 – System Setup and Basic Service Manual 7-45
4700-0043-00 Rev. 4
Common Error Codes for the Invenia ABUS 2.0
The Invenia ABUS 2.0 system error messages have been
improved to provide an easier troubleshooting experience for
field engineers and remote technical support team members.
The Invenia 2.0 system displays an error/information message
along with an error code. The error code displayed uses the
same error code rule (5 digit number) implemented with the
Invenia R1 system:
NOTE: The software team ensured that the new GEUS (Ultrasound
Box) error codes does not overlap with old Verasonics (VDAS/
Acquisition) codes. Thus, the error codes related to GEUS and
VDAS are unique to each system design.
Error Code (total 5 digits)
First 1 digit Second 1 digit Third 3 digits
1 - ABUSApp
related
1 - Acquisition related Acquisition error code
2 - Scanhead related 001 - Scanhead id does not match.
3 - Renderer related
001 - Image renderer initialization
failed.
4 - System status related
2 - ScannerApp
related
1 - Initialization related
ScannerApp initialization related error
code
2 - USB related
ScannerApp USB related error code
after initialization
3 - GEUS related
1 - Failure Type
GEUS fatal error information category
(requires system shutdown)
2 - Info Type
GEUS information error information
category (warning/information)
3 - Service Type
GEUS service error information
category (pass to service app)