Site Preparations
2-18 Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4
Environmental Dangers
Commercial devices such as laser cameras, printers, VCRs and
external monitors, usually exceed allowable leakage current
limits and, when plugged into separate AC outlets, are in
violation of patient safety standards. Suitable electrical isolation
of such external AC outlets, or providing the device with extra
protective earth, will be required in order to meet UL60601-1
and IEC60601-1 / IEC60601-1-1 standards for electrical
leakage.