General requirements
Invenia ABUS 2.0 – System Setup and Basic Service Manual 2-3
4700-0043-00 Rev. 4
Electrical requirements
General requirements
The Invenia ABUS 2.0 requires a dedicated power and ground
for the proper operation of its Ultrasound equipment. A separate
power outlet with a minimum 10 amp circuit breaker for
100-240 VAC.
NOTE: Please note that image artifacts can occur, if at any time within
the facility, the ground from the main facility's incoming power
source to the Invenia ABUS 2.0 is only a conduit.
Site power outlets
A dedicated AC power outlet must be within reach of the Invenia
ABUS 2.0 without extension cords. Other outlets adequate for
the external peripherals, medical and test equipment needed to
support this Invenia ABUS 2.0 must also be present within 1 m
(3.2 ft.) of the Invenia ABUS 2.0. Electrical installation must
meet all current local, state, and national electrical codes.
Unit power plug
Refer to Chapter 9 for Power Plug part numbers. If the Invenia
ABUS 2.0 arrives without a power plug, or with the wrong plug,
you must contact your GE dealer or the installation engineer
must supply what is locally required.
CAUTION
Power outage may occur. The Invenia ABUS 2.0 requires a
dedicated single branch circuit. To avoid circuit overload and
possible loss of critical care equipment, make sure you do not
have any other equipment operating on the same circuit.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Invenia ABUS 2.0 and is the answer not in the manual?

GE Invenia ABUS 2.0 Specifications

General IconGeneral
BrandGE
ModelInvenia ABUS 2.0
CategoryMedical Equipment
LanguageEnglish

Related product manuals