Components and Functions (Theory)
5-22 Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4
Stabilizing Membrane
The Scan Head senses whether the membrane is attached, and
if it is missing, the system alerts the operator with the error
message “No membrane is detected”. Also, a missing
membrane is reported in the Service Desk Diagnostics.
If a Membrane is not in place, the scan cannot begin until the
membrane has been properly mounted.
Handles and operator control pushbuttons
The Scan Head has a left and a right handle. These are used for
head positioning and for the operator to adjust compression
either manually or by pushbutton.
The left handle has two pushbuttons, one to start a scan and
one to abort a scan. The right handle has two pushbuttons, one
to increase compression and one to decrease compression.
Pressing a button once changes the weight by one step (about 7
pounds force). The touchscreen displays a picture identifying
compression state.