EasyManua.ls Logo

GE LOGIQ 5 - Section 4-9 Board Function Checks

GE LOGIQ 5
452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
D
IRECTION 2300000, REVISION 2 LOGIQ™5 SERVICE MANUAL
Chapter 4 Functional Checks 4-41
Section 4-9Board Function Checks
4-9-1 RLY Assy function check procedure
1.) Turn on the system
2.) After completion of echoloader loading, connect probe on port1. See “Connecting a probe” on
page 4-12.
3.) Log into "Service Login". Refer to Section 5-10-2 - Global Service User Interface (GSUI) on page
5-33.
4.) Click "Diagnostics" on top menu.
5.) On left folder window, select 'LOGIQ 5 Diagnostics'=>'FRU test'=>'RLY'.
6.) Select 'RLY Full Test' and execute test.
7.) All diagnostic test items should be passed. Refer to the figure below.
8.) Activate probe. Refer to Section 4-3-7-2 - Activating the probe on page 4-12.
9.) Check the system basic functions. Refer to Section 4-3-4 - System B/M-Mode Checks on page 4-
7. Refer to Section 4-3-5 - System CFM and PWD Checks on page 4-9.
10.)Disconnect probe. Refer to Section 4-3-7-4 - Disconnecting the probe on page 4-12.
11.)Connect probe on port2 and repeat step 8) ~ 10).
12.)Connect probe on port3 and repeat step 8) ~ 10).
13.)All basic functions must show correct operation as described.
Figure 4-42 RLY diagnostic test

Table of Contents