GE MEDICAL SYSTEMS
DIRECTION 2300000, REVISION 1 LOGIQ™ 5 SERVICE MANUAL
7 - 12 Section 7-7 - LOGIQ5 Diagnostics
Section 7-7LOGIQ5 Diagnostics
When you select a test to run, the display changes as shown in Figure 7-8 on page 7-13 .
1.) Instructions Frame
- Displays either a test-specific text or the default instructions.
2.) Status Frame
- Initially displays the last known status for a selected diagnostic. Once the diagnostic starts,
the frame displays the “current” status of all test results.
- The Status Frame also contains the user interface elements used for Diagnostic Control and
Operator Feedback.
3.) Execute Button
- This switch has two modes - each with appropriate text:
- Execute - to start the diagnostic
- Abort - to stop a diagnostic
4.) Loop Count
- This is an editable text field that will only accept numeric values of 4 digits or less. When the
switch is configured as an “execute” switch and pressed, the loop count field will be queried
to determine the number of times to execute the diagnostic.
5.) Progress Indicator
- Displays a graphical progress indicator for the user.
6.) Short Text Message
- Displays brief messages about the test’s progress during execution.
7.) Status Frame Background Color
- Initially gray, the Status Frame background color changes upon completion of a diagnostic to
indicate completion status.
- Code Status Fail = Red
- Code Status Pass = Green
- Code Status neither Pass nor Fail (Example: Aborted) = Gray (default color).