GE MEDICAL SYSTEMS
D
IRECTION 2300000, REVISION 2 LOGIQ™5 SERVICE MANUAL
Chapter 8 Replacement Procedures 8-91
8-7-1-4 Removal Procedure (cont’d)
7.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-7-1-5 Mounting Procedure
Install the new parts in the reverse order of removal.
Table 8-43 Functional Tests
Service Manual
Section Functional Test / Diagnostic Test Debrief Script
Section 4-3-1
Power On/Boot Up
“Service Manual, Direction
2300000, Rev 1+, Section 8-7-1. Equipment
passed all required tests and is ready for use. “
Section 4-3-2
Power Off / Shutdown
Section 4-9-1
RLY Assy function check procedure
Section 4-3-4
System B/M-Mode Checks
Section 4-3-5
System CFM and PWD Checks
Section 10-5-2
Functional Checks (See Also Chapter 4)
Section 10-6
Using a Phantom

Table of Contents