DIRECTION 2300000, REVISION 2 LOGIQ™5 SERVICE MANUAL
8-184 Section 8-11 - Installing Base Software Load Image
8-11-1-1 System Checks
1.) Approx. two minutes later, the windows XP desktop appears. Then one minutes later, the following
screen appears. Clisck on Yes using right button of the trackball to restart the system
NOTE: If the window above is not displayed, go to next step
2.) Drive Letter Checks:
a.) Double click on My Computer
b.) First check to see if you need to change the drive letters on the CDROM drive and the
Magneto Optical drives – they SHOULD
be G: and H: respectively.If they are, go to next
step 3.) Driver Checks
Figure 8-224 System Setting Change
Figure 8-225 My Computer