IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
3-98 Section 3-10 - Connectivity Setup
3-10-3 Connectivity Configuration
NOTE: If connected to a stand-alone network (Peer-to-Peer network with a Vivid q N scanner, an
EchoPAC PC work station and eventually a network printer), you should use default delivery
settings.
3-10-3-1 Introduction
To be able to use the network functions when connected to a hospital network, the scanner must have
a proper network address.
Before you can set up the scanner, you need to collect some information. Refer to the "Worksheet
for DICOM Network Information in Figure 2-2 on page 2-10. Typically, the source for this
information is the network administrator.
Follow the steps below to prepare the scanner for use on the network.
3-10-3-2 Select TCP/IP Set-up Screen
1.) Press CONFIG
(F2) and log on as Adm, as described in "Log On to the System as ADM" on page
4-10.
2.) If not already selected, select CONNECTIVITY
from the bottom of the screen.
3.) Select the TCP/IP TAB (it is named Tcpip).
.
Figure 3-100 TCP/IP Set-up Screen for Vivid q N, Overview (Example)
Computer Name:
For Vivid q N, this name is on the form:
Vivid_I-00nnnn, where “00nnnn” is a number
(nnnn is the scanner’s serial number).
IP Settings:
IP Settings area for the Vivid q N scanner.
Remote Archive Setup area:
IP settings area for a Remote Archive.
Example: EchoServer.
Default Setup:
Remote Archive IP-Addr: 10.0.0.4
Remote Archive Name: EchoPAC7-000001