IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
8-2 Section 8-2 - External Component Replacement Procedures
Section 8-2
External Component Replacement Procedures
8-2-1 Bearing Handle Replacement Procedure
NOTE: The term “Bearing Handle” relates to the Vivid q N portable carrying handle.
8-2-1-1 Tools
Use the appropriate flat and Phillips-type screw drivers as indicated in the Bearing Handle replacement
procedure.
8-2-1-2 Time Required
10 minutes
8-2-1-3 Preparations
None
8-2-1-4 Bearing Handle Removal Procedure
1) Close the Vivid q N LCD cover, making sure both front latches are securely fastened (Figure 8-1).
2.) Release the captive screw on either side of the Bearing Handle (there are two screws that secure
the handle in position), by turning 1/4 turn in a counterclockwise direction (Figure 8-2).
Figure 8-1 Closing Vivid q N LCD Cover
Figure 8-2 Releasing the Bearing Handle Captive Screw
LatchLatch
Captive Screw