GE
D
IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-55
8-3-8 AUO LCD Display Replacement Procedure
NOTE: The procedures described in this section are applicable to all Vivid q N BT06 systems with software
version 6.2.0 Build 158, and higher.
NOTE: The AOU LCD PN5461952 kit is delivered with the LCD monitor assembled in the display frame
as an integrated unit.
NOTE: If replacing an old-type LCD Monitor with an AUO LCD Monitor, it is necessary to uninstall the old-type
hinges and install a pair of new-type hinges (P/N KTZ280195), as shown in Figure 8-64.
NOTE: Prior to installing the AUO LCD monitor, a replacement DC power supply, P/N 5439596 should be
available if this is required. See Note on page 8-58.
8-3-8-1 Tools
Use the appropriate Phillips and flat screw drivers, and Loctite 402 glue, as indicated in the AUO LCD
Display replacement procedure.
8-3-8-2 Time Required
15 minutes
8-3-8-3 Preparation
1) Make sure all required tools and parts are available.
2.) Adhere the appropriate Vivid label onto the new display frame.
3.) Remove the display frame from the new AOU LCD Display (to provide access to the hinges).
4.) Shut down the Vivid q N ultrasound unit, as described in - System Setup
8-3-8-4 AUO LCD Display Removal Procedure
1) Remove the bearing handle, as described in the "Bearing Handle Removal Procedure" on page 8-2.
2) Remove the Control Panel and Keyboard as described in the "Control Panel and Keyboard
Removal Procedure" on page 8-10.
Figure 8-63 AUO LCD Display Assembled in Display Frame