EasyManua.ls Logo

GE Vivid q N - Confirming Safelock Cart Voltage Configuration; System Voltage Confirmation

GE Vivid q N
698 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE
D
IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
3-18 Section 3-5 - Preparing for Installation
3-5-4 Confirming SafeLock Cart Voltage Configuration
1.) Check the safety rating plate located at the rear of the SafeLock Cart, above AC input connector.
2.) Remove the bottom cover (rear) from the SafeLock Cart to provide access to the jumper on the AC
unit. See Figure 3-11 below.
3.) Make sure that the jumper is set to either 100 to 120 V AC or to 220 - 240 V AC, in accordance with
the voltage rating shown on the safety rating plate (step 1).
3-5-5 System Voltage Confirmation
3-5-5-1 System Voltage Settings
Verify that the Vivid q N is set to the correct voltage.
The VividqN voltage settings are found on the base (underside) of the system.
For illustrations of the label with the voltage setting, see:
Table 1-5 "Main Label for Vivid q N" on page 1-19
Figure 3-11 Jumper on AC Unit
WARNINGWARNING
CONNECTING THE VIVID Q N TO THE WRONG VOLTAGE LEVEL WILL MOST LIKELY
DESTROY THE SCANNER.

Table of Contents

Related product manuals