IRECTION FQ091019, REVISION 2 VIVID Q N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-47
8-3-4 BIOS Battery Replacement
NOTE: Since the BIOS battery is not available as a spare part (due to shelf life), it should be obtained
locally. The battery model is CR2032.
8-3-4-1 Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the BEP replacement procedures.
8-3-4-2 Time Required
15 minutes
8-3-4-3 Preparations
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
8-3-4-4 BIOS Battery Removal Procedure
1) Remove the BEP, as described in the "BEP Removal Procedure" on page 8-40.
2.) Take out the BIOS battery located on the bottom of the BEP, as shown in Figure 8-55.
8-3-4-5 BIOS Battery Installation Procedure
NOTE: Before proceeding, refer to the "Preparation Procedure" on page 8-45.
1) Fit a new BIOS battery in position on the BEP - refer to Figure 8-55.
2.) Re-install the BEP, as described in the "BEP Installation Procedure" on page 8-44.
3.) Turn ON the system and boot-up.
4.) Adjust to the correct date and time settings - see "Date and Time Adjustments" on page 3-71.
5.) Proceed to perform the following functionality test:
- "Battery Parameters Test" on page 7-102
Figure 8-55 BEP - Location of BIOS Battery