8-2-8 LCD Display Frame Replacement Procedure
8-2-8-1 Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the bottom assembly replacement
procedures.
8-2-8-2 Time Required
10 minutes
8-2-8-3 Preparations
Shut down the Vivid q N ultrasound unit, as described in Chapter 3 - System Setup.
8-2-8-4 LCD Display Frame Removal Procedure
1) With the LCD Display in the open (upright) position, use a flat screwdriver to remove the screw cap
from each of the four screws that fasten the LCD display frame in position, as shown in Figure 8-28
below.
Figure 8-28 Removing the Screw Caps from the LCD Display Frame
2.) Loosen and remove the four screws (Figure 8-29).
3.) With the LCD Display in the fully open position, carefully release the LCD display frame from the
Figure 8-29 Removing the Securing Screws